SOURCES SOUGHT NOTICE FOR PREVENTIVE MAINTENANCE SERVICES FOR EQUIPMENT STERILIZERS VA MEDICAL CENTER AT 10 N. GREENE ST., BALTIMORE, MD

Agency: VETERANS AFFAIRS, DEPARTMENT OF
State: Maryland
Type of Government: Federal
FSC Category:
  • J - Maintenance, Repair, and Rebuilding of Equipment
NAICS Category:
  • 811310 - Commercial and Industrial Machinery and Equipment (except Automotive and Electronic) Repair and Maintenance
Posted Date: Mar 16, 2025
Due Date: Mar 24, 2025
Solicitation No: 36C24525Q0382
Original Source: Please Login to View Page
Contact information: Please Login to View Page
Bid Documents: Please Login to View Page

Description

Follow
SOURCES SOUGHT NOTICE FOR PREVENTIVE MAINTENANCE SERVICES FOR EQUIPMENT STERILIZERS VA MEDICAL CENTER AT 10 N. GREENE ST., BALTIMORE, MD
Active
Contract Opportunity
Notice ID
36C24525Q0382
Related Notice
Department/Ind. Agency
VETERANS AFFAIRS, DEPARTMENT OF
Sub-tier
VETERANS AFFAIRS, DEPARTMENT OF
Office
245-NETWORK CONTRACT OFFICE 5 (36C245)
Looking for contract opportunity help?

APEX Accelerators are an official government contracting resource for small businesses. Find your local APEX Accelerator (opens in new window) for free government expertise related to contract opportunities.

APEX Accelerators are funded in part through a cooperative agreement with the Department of Defense.

The APEX Accelerators program was formerly known as the Procurement Technical Assistance Program (opens in new window) (PTAP).

General Information
  • Contract Opportunity Type: Sources Sought (Original)
  • Original Published Date: Mar 14, 2025 12:33 pm EDT
  • Original Response Date: Mar 24, 2025 12:00 pm EDT
  • Inactive Policy: 15 days after response date
  • Original Inactive Date: Apr 08, 2025
  • Initiative:
    • None
Classification
  • Original Set Aside:
  • Product Service Code: J065 - MAINT/REPAIR/REBUILD OF EQUIPMENT- MEDICAL, DENTAL, AND VETERINARY EQUIPMENT AND SUPPLIES
  • NAICS Code:
    • 811310 - Commercial and Industrial Machinery and Equipment (except Automotive and Electronic) Repair and Maintenance
  • Place of Performance:
    Baltimore , MD 21201
    USA
Description

DISCLAIMER



The sources sought notice is solely for information and planning purposes only and does not constitute a solicitation will be issued. NOTICE The sources sought notice does not constitute a Request for Quotation (RFQ) or a promise to issue an RFQ in the future. Responses to this notice are not offers and cannot be accepted by the Government to form a binding contract. Responders are advised that the Government will not pay for any information or administrative costs incurred in response to this notice. It is the responsibility of the responders to monitor this site for additional information pertaining to this requirement. Responses or questions concerning the composition and requirements to this notice will not be returned.



All capable contractors are welcome to respond to this notice and shall provide a brief capability statement that covers the information in the statement of work. The Capability Statement shall be limited to 5 pages. The Government will not review any other information or attachments that is more than the mentioned page limit. The Government will not review generic marketing materials that do not address the information contained herein or the attached documents.



GENERAL Your response must address capabilities specific to the services required in the statement of work and must include the following:



a. Company Name



b. SAM Unique Entity ID & cage code which the company is registered in sam.gov and https://veterans.certify.sba.gov/



c. Company Address



d. Point of contact(s) name



e. Telephone number(s)



f. Email address(s)



g. Socioeconomic data



h. Company Business Size Status and representations of your business



i. Any previous contract(s) similar or the same as the required functional capabilities in the statement of work



This is not a request for quotes. The estimated period of performance is September 2025 through August 2026. All interested parties shall respond no later than 3/24/2025 at 12:00PM EST. Submit responses or suggestions via email to Kenneth.dougherty@va.gov. Telephone responses will not be accepted.







STATEMENT OF WORK




  • 4 Preventive maintenance visits a year for a Beta Star steam sterilizer, model N243660, serial# B0337, to ensure compliance with FDA guidelines for animal housing.

  • As needed emergency service calls with 12 hour response time.

  • PM Inspection Checklist-




  • Visually check the outside for signs of leaks or damage

  • Check steam, water and air supply lines for leaks

  • Check the door travel operation

  • Check and adjust door for 1/8" clearance

  • Check the lower door switch operation

  • Check the door lock cylinder operation

  • Check the power door cylinder operation (if automatic)

  • Check for proper door seal retraction

  • Check the inside of the chamber for melted materials and clean

  • Pull and clean the drain strainer

  • Clean and check the chamber screen

  • Check the door gasket and replace if necessary

  • Check the traps, valves, solenoid and the plumbing for leaks

  • Check load cart or shelf rack for proper alignment

  • Check the airlines and valves for leaks

  • Check liquid cycle

  • Check the slow exhaust rate

  • Check the operation of the vacuum system

  • Check the 2 PSI pressure AND operation

  • Run solid cycle @ 121ºC

  • Run bio bag cycle @ 121ºC

  • a.) Check the prevac depth

  • b.) Check the charge pressure

  • Check "water in chamber" safety sensor operation

  • Check the dry cycle phase for proper operation

  • Perform a leak test (annually)

  • Replace air-in filter annually (indicate date changed)

  • Check tempering valve for proper effluent quench






  • The equipment is in the animal facility of the VA Hospital at 10 N Greene St, Baltimore, MD 21201.






  • All parts included except pumps/motors/controls





Location(s): Baltimore Veterans Affairs Medical Center, 10 North Greene Street, Baltimore, MD 21201





End


Attachments/Links
Contact Information
Contracting Office Address
  • 8849 International Drive Suite 215
  • LINTHICUM , MD 21090
  • USA
Primary Point of Contact
Secondary Point of Contact


History
  • Mar 14, 2025 12:33 pm EDTSources Sought (Original)
Get Government Bids Like This by Email Receive daily bid alerts that match your keywords, business categories, and target regions.

See Also

ID: BPM058517 Title: WOELCT30SB- Electrical Maintenance and Repairs- MDOT/SHA Worcester County Status: Open

State Government of Maryland

Due by 9/16/2026

ID: BPM058563 Title: Cannabis Incubator Facility Operator Status: Open Due / Close Date:

State Government of Maryland

Due by 9/30/2026

RFQ-2027-001 TRUCK SCALE, EQUIPMENT MAINTENANCE AND REPAIR Opening Date: September 16, 2026 Opening

Howard County

Due by 9/16/2026

Follow TRS-3D /AN/SPS-75 Radar repairs/ overhaul services Active Contract Opportunity Notice ID 70Z08526RRJ001B00

HOMELAND SECURITY, DEPARTMENT OF

Due by 12/01/2026

* Disclaimer: This website provides information about bids, requests for proposals (RFPs), or requests for qualifications (RFQs) for convenience only and does not serve as an official public notice. Individuals who wish to respond to or inquire about bids, RFPs, or RFQs should contact the relevant government department directly.