EAST BAY MUNICIPAL UTILITY DISTRICT
REQUEST FOR QUOTATION (RFQ) No. 2616
for
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER WITH
AUTOSAMPLER
Contact Person: John W. Grimes, Buyer II
Phone Number: (510) 287-0316
E-mail Address: john.grimes@ebmud.com
For complete information regarding this project, see RFQ posted at
https://www.ebmud.com/business-center/materials-and-supplies-bids/current-requests-
quotation-rfqs/ or contact the EBMUD representative listed above. Please note that prospective
bidders are responsible for reviewing this site during the RFQ process, for any published addenda
regarding this RFQ.
Bids Due
by
2:00 p.m.
on
March 18, 2026
All bid submissions hand delivered or mailed (USPS, FedEx, UPS, etc.) to the address or
PO Box noted below and must be received no later than 2:00 p.m. on the bid due date.
RESPONSE DELIVERED BY SERVICE
(UPS, FedEx, DHL, etc., during
business hours: 8:00 AM – 3:30
PM only) to:
RESPONSE DELIVERED BY MAIL
(U.S. Postal Service) to:
EBMUD–Purchasing Division
INDUCTIVELY COUPLED PLASMA
MASS SPECTROMETER WITH
AUTOSAMPLER – RFQ 2616
375 11th Street
Oakland, CA 94607
REV: 9/23/25
EBMUD–Purchasing Division
INDUCTIVELY COUPLED PLASMA
MASS SPECTROMETER WITH
AUTOSAMPLER – RFQ 2616
P.O. Box 24055
Oakland, CA 94623
RESPONSE HAND-DELIVERED
(during business hours: 8:00
AM – 4:00 PM only)
EBMUD–Purchasing Division
INDUCTIVELY COUPLED
PLASMA MASS SPECTROMETER
WITH AUTOSAMPLER – RFQ
2616
Purchasing Office
375-11TH Street, 1st Floor
Oakland, CA 94607
EAST BAY MUNICIPAL UTILITY DISTRICT
RFQ No. 2616
for
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER WITH AUTOSAMPLER
TABLE OF CONTENTS
I. STATEMENT OF WORK ........................................................................................................... 3
A. SCOPE................................................................................................................................... 3
B. BIDDER QUALIFICATIONS .................................................................................................... 3
C. SPECIFIC REQUIREMENTS .................................................................................................... 3
D. FAILURE TO MEET SPECIFICATIONS................................................................................... 16
II. CALENDAR OF EVENTS ......................................................................................................... 17
III. DISTRICT PROCEDURES, TERMS, AND CONDITIONS .............................................................. 17
A. RFQ ACCEPTANCE AND AWARD ........................................................................................ 17
B. BRAND NAMES, APPROVED EQUIVALENTS, DEVIATIONS, AND EXCEPTIONS .................. 17
C. PRICING.............................................................................................................................. 18
D. NOTICE OF INTENT TO AWARD AND PROTESTS................................................................ 18
E. METHOD OF ORDERING .................................................................................................... 19
F. TERM / TERMINATION / RENEWAL ................................................................................... 20
G. WARRANTY ........................................................................................................................ 20
H. INVOICING ......................................................................................................................... 20
IV. RFQ RESPONSE SUBMITTAL INSTRUCTIONS AND INFORMATION ......................................... 21
A. DISTRICT CONTACTS .......................................................................................................... 21
B. SUBMITTAL OF RFQ RESPONSE ......................................................................................... 21
ATTACHMENTS
EXHIBIT A – RFQ RESPONSE PACKET
EXHIBIT B – INSURANCE REQUIREMENTS
EXHIBIT C – GENERAL REQUIREMENTS
REV: 9/23/25
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER WITH AUTOSAMPLER
I. STATEMENT OF WORK
A. SCOPE
It is the intent of these specifications, terms, and conditions to describe requirement to
furnish, install, test and validate one (1) new bench top inductively coupled plasma mass
spectrometer with autosampler f.o.b. destination to the District’s SD-1 Laboratory
located at 2020 Wake Avenue, Oakland, CA 94607.
East Bay Municipal Utility District (District) intends to award a contract to the lowest
cost bidder(s) whose response meets the District’s requirements.
B. BIDDER QUALIFICATIONS
1. Bidder Minimum Qualifications
a. Bidder, bidder’s principal, or bidder’s staff shall have been regularly
engaged in the business of providing inductively coupled plasma mass
spectrometers for at least ten (10) years.
b. Bidder shall be a certified and/or authorized manufacturer and dealer of
these instruments.
c. Bidder shall possess all permits, licenses, and professional credentials
necessary to supply product and perform services as specified under this
RFQ.
C. SPECIFIC REQUIREMENTS
All products shall be in new and unused condition and shall be of the most current and
up to date model.
1. GENERAL – QUADROPOLE ICP-MS INSTRUMENT
The vendor shall furnish, install, test and validate one new bench top inductively
coupled plasma mass spectrometer (ICP-MS) for trace metals analysis. The ICP-
MS must satisfy the specifications outlined in the Criteria Table. The instrument
will detect metals in source water, ground water, drinking water, wastewater,
salt water and industrial waste for analysis at concentrations ranging from 0.4
ppb to 1000 ppb. The instrument acceptance process shall include calibration of
analytes of interest in the Table for EPA 200.8. The method verification will
include running standards at the required method reporting limits for drinking
water and wastewater compliance, demonstrating instrument check blanks
RFQ No. 2616
Page 3
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER WITH AUTOSAMPLER
below the Method Detection Limits (MDLs) in the Table and provide an Initial
Demonstration of Capability (IDOC) that meets all method quality control criteria.
The system must include an enclosed autosampler with connection to ventilation,
a liquid sample introduction system, an RF argon plasma ion source, a
quadrupole based cell to eliminate spectroscopic interferences with KED using
collision gases, quadrupole mass filter, simultaneous dual stage discrete dynode
detector and at least two-cone plasma interface, all under computer control.
Note: EBMUD will provide an I7 WIN11 PC with 32gb RAM that the bidder will
need to work with District IT to install the software on site during instrument
installation.
The table below is divided into sections of instrument criteria. For the ‘Location in Proposal’ field, please
indicate where in the proposal the information is provided that indicates this criteria is met (a page # and
paragraph or section). For those items that are not shaded in the last field, please also provide a reference to
an attachment of raw data or other instrument documentation that indicates the item criteria is met and a
customer reference that currently uses the instrument in this configuration. For example, Section 2b requires
documentation and a reference to confirm MDLs have been met. Please provide the raw data and a summary
of tabulated MDL data from the instrument with a date and time stamp in a spreadsheet file. Please use a
minimum amount of documentation to confirm availability of these criteria, such as a single page.
Section
2
a
b
c
d
e
f
g
h
i
j
k
Item
ICP-MS GENERAL SYSTEM REQUIREMENTS
System must be fully compliant with US EPA methodology for Method
200.8
Based on analyzing MB & low-level spikes using the calibration in Table
1, the system must meet or exceed the MDLs in Table 1 for both Water
& Wastewater
Fully automated system startup, shutdown and optimization and
computer-controlled instrument tuning for optimum performance. This
includes software startup & shutdown of equipment used for cooling the
system, such as a chiller or other heat exchange device
Quadrupole must be able to scan a minimum of 3000 amu/sec to
achieve maximum productivity from transient signals when jumping
between light masses and high masses
Quadrupole resolution settings are dynamically scanned as each element
is analyzed
Mass range must be at least 4 to 260 amu or greater
Remote service trouble shooting and remote operation of system
All sample introduction, analytical system, data acquisition, method
validation and quality control operations using a single software
program
Full system must use single-phase power and remain in specification
temperature range of 10 to 30°C at up to 2°C per hour
System must be able to be sufficiently cooled in laboratory temperatures
below 30°C
Software automation available to optimize operating conditions and
pass tune with each batch
Location in Proposal
Data Included
RFQ No. 2616
Page 4
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER WITH AUTOSAMPLER
Section
l
3
a
b
c
d
e
f
g
h
Item
System must be able to calibrate, detect and use correction equation for
1/2 masses and detection of Nd, Sm, Gd, Dy
SAMPLE INTRODUCTION
Sample introduction system has concentric nebulizer, cyclonic or double
pass spray chamber, quartz injector and no o-rings
Sample introduction uses a computer controlled integral peristaltic
pump. Must be integrated within the ICP-MS system
System is configured for automated addition of internal standards to all
samples and standards
Entire torch and injector assembly can be removed and installed by hand
in a tool-less fashion
Cone and torch apparatus single orientation or one fit-type for ease of
assembly
Fully automated one touch X, Y, Z torch alignment
No carryover >MDL to subsequent samples, standards and blanks from
samples or standards with concentrations up to 1000 ppb
Linear Dynamic Range up to 1000 ppb
Location in Proposal
Data Included
i
Sample introduction system is capable of providing an aerosol as well as
liquid dilution.
4
ION SOURCE AND RF GENERATOR
a
Computer controlled RF generator allows for automatic control of torch
ignition, shutdown, and system warm up
b
Allows independent adjustment of the ion optic parameters and the
plasma conditions
5
PLASMA INTERFACE
a
Design with rapid mounting and removal cone design, easily accessible
from the front of the instrument
Mechanism for complete separation of ions from photons and neutrals
b
either with 90 degree bend of ion path or a chicane , making the cell and
mass analyzing quadrupole completely maintenance free
6
COLLISION CELL
System must incorporate a collision cell with two modes of operation:
a
Standard Mode and Collision Cell Mode with KED and the ability to
utilize a wide variety of gases for the life of instrument with no
degradation to the gas channels, turbo pump or cell
In Standard Mode, the collision cell must be able to operate with no gas
b
and no energy barrier for optimal signal-to-noise performance on
elements with any interference
In Collision Cell Mode, the collision cell must be able to operate as a
c
passive ion guide with an energy barrier at the cell exit for KED. The
flow and optimization of the gases shall be under full operator control
d
Instrument provides transition into and out of collision cell mode and
complete stability in <3 seconds
7
QUADRUPOLE MASS ANALYZER
RFQ No. 2616
Page 5
This page summarizes the opportunity, including an overview and a preview of the attached documents.
* Disclaimer: This website provides information about bids, requests for proposals (RFPs), or requests for qualifications (RFQs) for convenience only and does not serve as an official public notice. Individuals who wish to respond to or inquire about bids, RFPs, or RFQs should contact the relevant government department directly.

Sign-up for a Free Trial, Government Bid Alerts

With Free Trial, you can:

You will have a full access to bids, website, and receive daily bid report via email and web.

Try One Week FREE Now

See Also

Project ID: 2026-IFB-LM-0046 Title: Replace CAISO Remote Intelligent Gateway (RIG) and CAISO Meters

City of Pasadena

Bid Due: 6/08/2026

General Information Due Date: Thursday, June 18, 2026 Opening Time: 2:00 PM Published

City of Stockton

Bid Due: 6/18/2026

General Information Due Date: Thursday, June 18, 2026 Opening Time: 2:00 PM Published

City of Stockton

Bid Due: 6/18/2026

Bid Number: B-130000883 Bid Title: Two (2) Chevrolet Silverado 2500 or Equivalent Category:

City of Ventura

Bid Due: 6/17/2026