Request For Quote - DART-Mass Spectrometer
| Agency: | HOMELAND SECURITY, DEPARTMENT OF |
|---|---|
| State: | Virginia |
| Type of Government: | Federal |
| FSC Category: |
|
| NAICS Category: |
|
| Set Aside: | No Set aside used |
| Posted Date: | Sep 14, 2026 |
| Due Date: | Sep 23, 2026 |
| Solicitation No: | 192126VHQFDL00016 |
| Original Source: | Please Login to View Page |
| Contact information: | Please Login to View Page |
| Bid Documents: | Please Login to View Page |
Description
APEX Accelerators are an official government contracting resource for small businesses. Find your local APEX Accelerator (opens in new window) for free government expertise related to contract opportunities.
APEX Accelerators are funded in part through a cooperative agreement with the Department of Defense.
The APEX Accelerators program was formerly known as the Procurement Technical Assistance Program (opens in new window) (PTAP).
- Contract Opportunity Type: Combined Synopsis/Solicitation (Updated)
- Updated Published Date: Sep 14, 2026 02:18 pm EDT
- Original Published Date: Sep 14, 2026 08:59 am EDT
- Updated Date Offers Due: Sep 23, 2026 08:00 am EDT
- Original Date Offers Due: Sep 23, 2026 08:00 am EDT
- Inactive Policy: Manual
- Updated Inactive Date: Sep 24, 2026
- Original Inactive Date: Sep 24, 2026
-
Initiative:
- None
- Original Set Aside:
- Updated Set Aside: No Set aside used
- Product Service Code: 6640 - LABORATORY EQUIPMENT AND SUPPLIES
-
NAICS Code:
- 334516 - Analytical Laboratory Instrument Manufacturing
-
Place of Performance:
VA 22102USA
Request for Quote (RFQ) – 192126VHQFDL00016 DART-Mass Spectrometer
1. Point of Contact: Chris Jackson, 8222 N. Belt Line Road, Irving, Texas 75063
Questions: All questins shall be submitted no later than 1:00 pm, Eastern Time, Thursday, September 17, 2026.
Email quotes to: Christopher.L.Jackson@ice.dhs.gov by 0800, Eastern, September 23, 2026.
Basis for Award:
The Government intends to evaluate quotes in response to this RFQ using the Lowest Price Technically Acceptable (LPTA) process. Quotes shall be evaluated in order of total price (lowest price first) until a technically acceptable quote is determined. Quotes will be evaluated for acceptability only. A quote shall be determined to be technically acceptable or unacceptable based on the rating definitions in the table below. The technically acceptable quote will then have its price evaluated for reasonableness. All quotes may not be evaluated at the Government’s discretion.
Rating Definitions:
Acceptable
The quotation meets and/or exceed the minimim specifications as listed below.
Unacceptable
The quotation does not meet the minimum specifications as listed below. .
2. Set Aside: None
3. NAICS: 334516 PSC: 6640
4. Description:
The Government would like to purchase a Direct Analysis in Real-Time Mass Spectrometer (DART-MS) that will fulfill the HSI Forensics Laboratory mission to perform rapid, non-destructive mass spectral analyses of inks, toners, dyes, polymers, papers, adhesives, and other materials related to travel, identity, and security documents.The government’s purchase would also include the following: Delivery, Setup/Installation, and Warranty Period of no-less than 1 year.
The DART source must be capable of rapid qualitative and quantitative analysis with no sample preparation of a wide range of polar and nonpolar materials including polymers and polymer additives, fuel additives, drugs and metabolites (either purified or in body fluids or tissues), explosives, foods and beverages, environmental contaminants, pesticides, and chemical warfare agents. The DART system must be controlled through the software provided. The DART must be a fully integrated (with the MS and software) and be a commercially available instrument that has previously been sold and installed into US Federal and State government laboratories, industrial and pharmaceutical companies, as well as universities and colleges
The DART-MS equipment must meet the following minimum requirements listed below:
-TAA compliant
- Must include DART JumpShot: DART JumpShot with Pulse gas flow is an easy way to complete ambient pressure desorption ionization mass spectrometry. JumpShot technology improves speed and throughput, reduces ambient background and allows for 95% helium saving over continuous flow DART.
All hardware required for operation of the DART-MS instrument – PC (and related peripherals that meets or exceeds the following specifications:
- Intel Core i5-13600 with Windows 11™ Pro or Mac OSX 64-bit Operating System,
- Minimum of 16GB RAM, 256GB SSD & 4TB SATA disk drives,
- 24" flat panel monitor
- optical mouse, keyboard, and all necessary cables
Includes accessory pumps, gas regulator(s), etc.
The Mass spectrometer must meet or exceed the following specifications:
- Time-of-flight mass spectrometer including ESI ion source, DART ion source, isolation valve, ion transmission guide, ion accelerator, analyzer, ion detection unit, ADC system, evacuation and control system.
- Syringe pump for system tuning.
- Positive and Negative ion detection
- No special interface or pump required for using DART
- Resolving Power: greater than or equal to 10,000 (FWHM) at m/z 609
- Data Acquisition Rate: up to 50Hz
- Mass Range of at least: m/z 7-10,000
- Mass Accuracy: 3ppm (RMS) or better
- Supports up to multiple in-source CID voltages for a single analysis
- Vacuum isolation valve permits rapid pump down after venting the ion source region for maintenance
- Must be capable of using Helium and Nitrogen gas as a carrier gas
- High-Resolution, capable of exact mass and true isotopic pattern measurements
- Data from analyses can be matched against database spectra measured from other models of DART-MS, including but not limited to the NIST DART Forensic Database, and the US Fish and Wildlife Forensic Laboratory wood database.
- Atmospheric pressure interface is user serviceable.
- Is a non-contact source that does not require that the sample be prepared in solvent or reside in a vacuum. The DART source can analyze vapors, solids, and liquid sample types
- MS is fully integrated with the direct analysis in real time (DART) source.
- A single software controls both the DART sample introduction and MS detector.
- The DART source does not use radiation and does not expose the sample to high voltages
- Capable of detecting masses as low as 7Li+ and atmospheric ions from m/z 15 and higher.
- Meets or exceeds spectral storage rate of at least 40 spectra per second at a resolving power in excess of 10,000 (FWHM definition).
- Can be shutdown and restarted routinely without detrimental effects to the instrument performance.
- can measure molecular ions, even for difficult nonpolar compounds such as saturated alkanes.
- Supports the use of optional ion sources such as atmospheric pressure ionization (APCI), nanoelectrospray (nanoESI), CoronaSpray, PaperSpray® and DualSpray.
-
Data interpretation and reporting software is provided that includes:
- elemental composition determination functions
- integrated searching of NIST-formatted mass spectral databases and on-line chemical databases for compounds with elemental compositions matching those calculated by the software.
- chemometric analysis should include principal component analysis (PCA) and kernel PCA, linear and kernel discriminant analysis (LDA and KDA), partial least squares (PLS), support vector machine (SVM), Fisher ratio analysis and heat maps. Heat maps can be exported to third-party software for hierarchical cluster analysis.
- Kendrick mass defect analysis including nontraditional and 3D Kendrick mass defect plots
- target compound identification and report generation
- support for creation of custom mass spectral databases/libraries
- modeling of overlapping isotopic distributions at any arbitrary resolving power
- Ion series analysis
- database searching including correlation and matching of mass spectra acquired with different collision energies with post-search exact-mass and elemental composition filters.
- The software is linked to databases such as NIST and Wiley mass spectral databases.
- fully off-line. Does not require internet/web connectivity for use of any function – database searching, data collection, analysis, or interpretation.
- can determine unique elemental compositions for unknown compounds by using isotopic exact masses and abundances. Provides information to support assignments for known or suspected compounds. The software to perform this assignment is provided.
- The government’s purchase would also include the following: Delivery, Setup/Installation, and Warranty Period of no-less than 1 year.
- 8222 N. BELT LINE ROAD SUITE 2200
- Irving , TX 75063
- USA
- Chris Jackson
- Christopher.L.Jackson@ice.dhs.gov
- Sep 14, 2026 02:18 pm EDTCombined Synopsis/Solicitation (Updated)
- Sep 14, 2026 10:54 am EDT Combined Synopsis/Solicitation (Updated)
- Sep 14, 2026 09:34 am EDT Combined Synopsis/Solicitation (Updated)
- Sep 14, 2026 08:59 am EDT Combined Synopsis/Solicitation (Original)
Related Document
| Sep 16, 2026 | [Combined Synopsis/Solicitation (Updated)] Request For Quote - DART-Mass Spectrometer |
See Also
IFB 161318 Box Culvert - Rt 611 Status: Open IFB 130851 Virginia Department
State Government of Virginia
Due by 11/19/2026
Follow NSN 6685016982840, SENSOR, AMBIENT TEMP, WSDC: JDN, ARLEIGH BURKE CLASS DDG Active
DEPT OF DEFENSE
Due by 11/09/2026
Follow Meter, Time Totalizing, [WSDC: 40A] End Item: Helicopter, UH-60A, UH-60L, UH-60V PR:
DEPT OF DEFENSE
Due by 11/19/2026
Follow Sextant, Marine; [WSDC: 23N] End Item: Ship, Ohio Class SSN (Trident) PR:
DEPT OF DEFENSE
Due by 11/20/2026