Onsite Preventative Maintenance and Calibration of Pipettes
| Agency: | HEALTH AND HUMAN SERVICES, DEPARTMENT OF |
|---|---|
| State: | Federal |
| Type of Government: | Federal |
| FSC Category: |
|
| Set Aside: | Total Small Business Set-Aside (FAR 19.5) |
| Posted Date: | Jun 9, 2025 |
| Due Date: | Jun 10, 2025 |
| Solicitation No: | 75F40125Q127670 |
| Original Source: | Please Login to View Page |
| Contact information: | Please Login to View Page |
| Bid Documents: | Please Login to View Page |
Description
APEX Accelerators are an official government contracting resource for small businesses. Find your local APEX Accelerator (opens in new window) for free government expertise related to contract opportunities.
APEX Accelerators are funded in part through a cooperative agreement with the Department of Defense.
The APEX Accelerators program was formerly known as the Procurement Technical Assistance Program (opens in new window) (PTAP).
- Contract Opportunity Type: Combined Synopsis/Solicitation (Updated)
- Updated Published Date: Jun 09, 2025 01:01 pm CDT
- Original Published Date: Jun 03, 2025 07:06 am CDT
- Updated Date Offers Due: Jun 10, 2025 10:00 am CDT
- Original Date Offers Due: Jun 10, 2025 10:00 am CDT
- Inactive Policy: 15 days after date offers due
- Updated Inactive Date: Jun 25, 2025
- Original Inactive Date: Jun 25, 2025
-
Initiative:
- None
- Original Set Aside: Total Small Business Set-Aside (FAR 19.5)
- Product Service Code: J066 - MAINT/REPAIR/REBUILD OF EQUIPMENT- INSTRUMENTS AND LABORATORY EQUIPMENT
-
NAICS Code:
- 811210 - Electronic and Precision Equipment Repair and Maintenance
-
Place of Performance:
Update: See below, questions that were submitted, and the Government's response thereto. The closing date of the solicitation is not extended.
Question 1: Can we provide 4 As Found readings and 10 As Left?
Response 1: No. The number of readings shall be same for “as found” and “as left” condition.
---------------------------------------------------------------------------------------------------------------------------
Question 2: How many samples are required for the Multi Channel pipettes?
Response 2: The question is not clear. Samples are not used for calibration. Water as a standard tested at three levels covering the pipette’s usable range (10%, 50%, 90%) for the calibration of a channel.
---------------------------------------------------------------------------------------------------------------------------
Question 2 part 2: How many readings need to be recorded for multi channel pipettes? Do they want 10 readings per channel at low, mid and high?
Response 2 part 2: 10 readings are required per channel at low, mid and high levels at both in “as found” and “as left” conditions.
---------------------------------------------------------------------------------------------------------------------------
Question 3: Are we allowed to use ISO8655 Tables 1 & 2 instead of the ones outlined in the solicitation? I have attached ISO8655 tables 1 & 2 for you.
Response 3: The calibration shall be done by the method provided.
---------------------------------------------------------------------------------------------------------------------------
Question 4: Can you confirm the start date in June and can we go into July to finish calibrations?
Response 4: Depending on contract award date, calibrations will be expected to begin in June; calibrations may go into July if necessary.
---------------------------------------------------------------------------------------------------------------------------
Question 5: When will the awarded company be notified?
Response 5: Upon receipt of award.
---------------------------------------------------------------------------------------------------------------------------
Question 6: Can the certificates of calibration be provided by noon the next day?
Response 6: Certificates shall be provided on the day of calibration; For pipettes calibrated close to the end of the day, certificates may be provided by noon on the following day.
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Original Post: See attached solicitation for Onsite Preventative Maintenance and Calibration of Pipettes, along with Attachment 1 - Pipette list, for US FDA's CBER at Silver Springs MD.
- 4041 Powder Mill Road 4th Floor
- Beltsville , MD 20705
- USA
- Suzanne Martella
- suzanne.martella@fda.hhs.gov
- Phone Number 8705437540
- Fax Number 8705437990
- Jun 09, 2025 01:01 pm CDTCombined Synopsis/Solicitation (Updated)
- Jun 03, 2025 07:06 am CDT Combined Synopsis/Solicitation (Original)
Related Document
| Jun 3, 2025 | [Combined Synopsis/Solicitation (Original)] Onsite Preventative Maintenance and Calibration of Pipettes |
See Also
Follow USNS LARAMIE (T-AO 203) MTA Active Contract Opportunity Notice ID N3220526R6058 Related
DEPT OF DEFENSE
Due by 9/24/2026
Follow BORESIGHT KIT CALIBRATION and REPAIR Active Contract Opportunity Notice ID N6833526R1106 Related
DEPT OF DEFENSE
Due by 9/24/2026
Follow T-6/T-34 Aircraft Maintenance and Contractor Logistics Support (CLS): Request for Proposal (RFP)
DEPT OF DEFENSE
Due by 10/08/2026
Follow LD5 Anchor Bar Refurbishment Active Contract Opportunity Notice ID W912ES26QA100 Related Notice
DEPT OF DEFENSE
Due by 9/25/2026