6515--Non-Invasive Ventilators
| Agency: | VETERANS AFFAIRS, DEPARTMENT OF |
|---|---|
| State: | Federal |
| Type of Government: | Federal |
| FSC Category: |
|
| NAICS Category: |
|
| Posted Date: | Apr 3, 2026 |
| Due Date: | Apr 9, 2026 |
| Solicitation No: | 36C24726Q0435 |
| Original Source: | Please Login to View Page |
| Contact information: | Please Login to View Page |
| Bid Documents: | Please Login to View Page |
Description
APEX Accelerators are an official government contracting resource for small businesses. Find your local APEX Accelerator (opens in new window) for free government expertise related to contract opportunities.
APEX Accelerators are funded in part through a cooperative agreement with the Department of Defense.
The APEX Accelerators program was formerly known as the Procurement Technical Assistance Program (opens in new window) (PTAP).
- Contract Opportunity Type: Sources Sought (Original)
- Original Published Date: Apr 03, 2026 04:13 pm EDT
- Original Response Date: Apr 09, 2026 04:30 pm EDT
- Inactive Policy: Manual
- Original Inactive Date: May 09, 2026
-
Initiative:
- None
- Original Set Aside:
- Product Service Code: 6515 - MEDICAL AND SURGICAL INSTRUMENTS, EQUIPMENT, AND SUPPLIES
-
NAICS Code:
- 339112 - Surgical and Medical Instrument Manufacturing
-
Place of Performance:
Department of Veterans Affairs Birmingham VAMC Birmingham , 35233-1927USA
This is NOT a solicitation announcement. This is a sources sought synopsis only. This synopsis aims to gain knowledge of potential qualified sources and their size classification relative to NAICS 339112 (size standard of 1,000 employees). Responses to this synopsis will be used by the Government to make appropriate acquisition decisions. After a review of the responses to this source sought synopsis, a solicitation announcement may be published.
The Department of Veterans Affairs (VA), Regional Procurement Office-RPO East Contracting Office, is seeking sources that can provide Non-Invasive Ventilation System for the Birmingham VA Healthcare System.
The Birmingham VAMC has a requirement for 12 Non-Invasive Ventilators, to include accessories, and staff training. Ventilators should meet the following salient characteristics:
1. Patient Population & Ventilation Range
Must support adult and pediatric patients 12.5 kg.
Must be capable of delivering tidal volumes down to approximately 100 mL.
2. Modes of Ventilation
The ventilator shall provide the following functional therapy modes:
Non Invasive Ventilation (NIV) with pressure support and backup respiratory rate (e.g., Spont PS, S/T or equivalent modes).
Invasive Pressure-Controlled ventilation.
Pressure regulated volume targeted ventilation (PRVC type functionality).
Proportional pressure support or equivalent capabilities to match patient effort.
Integrated high flow oxygen therapy capability.
3. Monitoring Capabilities
Must provide continuous CO and SpO monitoring, including during non invasive ventilation.
Monitoring capabilities must remain available during standby to support patient readiness.
Must store and display at least 72 hours of patient waveforms, parameters, and alarm history.
4. Portability & Power Requirements
Turbine driven ventilator (no external compressor required).
Weight shall not exceed approximately 17 lbs without trolley.
Must include:
Internal backup battery, and
Hot swappable external battery, providing combined runtime of at least 4 hours under typical conditions.
Must be suitable for intrahospital transport while maintaining ventilation.
5. Infection Control Features
Must include dual HEPA filtration or equivalent capability:
One filter protecting external air intake.
One filter on the patient gas pathway/exhalation side.
6. User Interface & Display
Minimum 12 inch color touchscreen or equivalent interface.
Must provide intuitive graphical monitoring and indicators for alarms and patient status.
7. Connectivity
Must support connectivity to common patient monitoring or EMR integration middleware (no brand required). Examples include:
Bedside monitor interface (e.g., via HL7, serial, or proprietary cable systems)
EMR middleware compatibility
8. Regulatory Requirement
Must be a FDA cleared Class II medical device meeting all applicable ventilation safety standards.
9. Accessories (Minimum Required)
Non invasive ventilation masks or interface.
Breathing circuits compatible with both invasive and non invasive modes.
Capnography sensor/interface suitable for NIV use.
Power supply/charger and mounting hardware.
The information identified above is intended to be descriptive, not restrictive, and to indicate the quality of the supplies/services will be satisfactory. It is the responsibility of the interested source to demonstrate to the government that the interested parties can provide the supplies/services that fulfill the required specifications. If you are interested and are capable of providing the sought-out supplies/services, please provide the requested information as well as the information indicated below.
Responses to this notice shall be submitted via email to darius.crane@va.gov no later than 16:30 EST. April 09, 202. Please include in your response company name, address, point of contact, and size of business. Include in the subject line of your email 36C24726Q0435 NON-INVASIVE VENTILATORS . Telephone responses shall not be accepted.
If a solicitation is issued it shall be announced later, and all interested parties must respond to that solicitation announcement separately from the responses to this source sought.
- ONE FREEDOM WAY
- AUGUSTA , GA 30904
- USA
- Darius Crane
- darius.crane@va.gov
- Phone Number (843) 577-5011
- Apr 03, 2026 04:13 pm EDTSources Sought (Original)
Related Document
| Apr 22, 2026 | [Combined Synopsis/Solicitation (Original)] 6515--NON-INVASIVE PULMONARY VENTILATORS |
See Also
Follow Dapagliflozin Tablets Active Contract Opportunity Notice ID SPE2D2-26-R-0018 Related Notice Department/Ind. Agency
DEPT OF DEFENSE
Follow Progesterone Capsules Active Contract Opportunity Notice ID SPE2D2-26-R-0015 Related Notice SPE2D2-26-R-0015 Department/Ind.
DEPT OF DEFENSE
Due by 9/18/2026
Follow 65--SURVIVAL KIT,INDIVIDUA Active Contract Opportunity Notice ID SPE2DS26T414R Related Notice Department/Ind. Agency
DEPT OF DEFENSE
Due by 9/15/2026
Follow 6505--Ticagrelor Tablets Active Contract Opportunity Notice ID 36E79726R0028 Related Notice 36E79726R0028 Department/Ind.
VETERANS AFFAIRS, DEPARTMENT OF
Due by 9/17/2026