J065--Siemens Symbia T16 Service Agreement John D. Dingell VA Medical Center
| Agency: | VETERANS AFFAIRS, DEPARTMENT OF |
|---|---|
| State: | Virginia |
| Type of Government: | Federal |
| FSC Category: |
|
| Posted Date: | Nov 22, 2024 |
| Due Date: | Nov 29, 2024 |
| Solicitation No: | 36C25025Q0143 |
| Original Source: | Please Login to View Page |
| Contact information: | Please Login to View Page |
| Bid Documents: | Please Login to View Page |
Description
APEX Accelerators are an official government contracting resource for small businesses. Find your local APEX Accelerator (opens in new window) for free government expertise related to contract opportunities.
APEX Accelerators are funded in part through a cooperative agreement with the Department of Defense.
The APEX Accelerators program was formerly known as the Procurement Technical Assistance Program (opens in new window) (PTAP).
- Contract Opportunity Type: Sources Sought (Original)
- Original Published Date: Nov 22, 2024 12:38 am EST
- Original Response Date: Nov 29, 2024 03:00 pm EST
- Inactive Policy: Manual
- Original Inactive Date: Jan 28, 2025
-
Initiative:
- None
- Original Set Aside:
- Product Service Code: J065 - MAINT/REPAIR/REBUILD OF EQUIPMENT- MEDICAL, DENTAL, AND VETERINARY EQUIPMENT AND SUPPLIES
-
NAICS Code:
- 811210 - Electronic and Precision Equipment Repair and Maintenance
-
Place of Performance:
John D. Dingell VA Medical Center Detroit , 48201USA
1. SCOPE OF WORK:
The Contractor shall provide all parts, labor, travel and technical phone support necessary to perform service and emergency repair on the Siemens Symbia T16 equipment located at the John Dingell VA Medical Center (VAMC) in strict accordance with all schedules, specifications, terms, conditions and provisions.
The Coverage plan includes:
Full services agreement-Siemens Symbia T16 (SN:2009)
Parts, Labor and Travel
Glassware Included
Crystals Included
Sources Included
Source Rod Included
Preventative Maintenance:
Two (2) Preventative Maintenance Visits per Year
To be performed during normal business hours (Mon-Fri, 8am-5pm) excluding federal holidays
2. PLACE OF PERFORMANCE:
John D. Dingell VA Medical Center
4646 John R Street, Detroit, MI 48201
3. CONFORMANCE STANDARD:
All services provided under this contract must be performed in conformance with the National Fire Protection Agency (NFPA), Occupational Safety and Health Administration (OSHA), and Original Equipment Manufacturer standards and specifications. The C&A requirements do not apply, and a Security Accreditation Package is not required.
4. HOURS OF WORK
Technical and Application Phone support is Monday through Friday from 8:00 a.m. to 5:00 p.m. EST
Onsite routine and PM services are 8:00 am through 5:00 pm EST, excluding federal holidays or as otherwise arranged with the Contracting Officer Representative (COR).
Exceptions are made for holidays. The eleven holidays observed by the Federal Government are New Year s Day, Martin Luther King Day, President s Day, Memorial Day, Juneteenth, Independence Day, Labor Day, Columbus Day, Veterans Day, Thanksgiving Day, and Christmas Day. Also, any other day declared by the President of the United States to be a national holiday.
5. PERFORMANCE REQUIREMENTS:
Contract shall include all normal spare parts, supplies necessary for repair and maintenance. Contractor shall have access and readily available manufacturer's replacement parts.
Equipment placed under this service contract shall be in good operating condition.
Preventive maintenance shall be done during normal working hours, Monday through Friday from 8:00 a.m. to 5:00 p.m., excluding federal holidays.
Scheduled preventive maintenance cycles shall be completed within one working day; extensions beyond one working day require the approval of the COR or Biomedical Engineering department designate.
Preventive maintenance schedule (PMI) shall be according to OEM specifications. This schedule shall be provided before the contract begins.
Unlimited emergency repair service shall be available on site, when necessary, Monday-Friday.
Performance verification shall be required after every preventive maintenance and emergency repair event. If any procedures were performed that would in any way affect the calibration status of the instrument, the results of performance testing shall conform to established performance criteria for the instrument.
Each maintenance or emergency repair event shall include inspection for internal and/or external causes of medical and electronic problems, including:
Wear, misalignment, maladjustment, damage or other malfunctions with respect to the manufacturer specifications.
Inspection for electrical safety and fire hazards.
Cleaning and lubrication of equipment as required.
Repair or replacement of all defective components necessary to conform to instrument specifications.
Calibration or standardization of instruments required.
Testing for proper operation of entire system and subsections.
Document proper function and performance of instrument.
Only new standard parts or reconditioned parts shall be furnished by the Contractor in effecting repairs. Parts supplied by the Contractor shall be installed by the Contractor.
All parts replaced within plan type coverage shall be identified by part number as shown on the schematics/parts list. Actual cost of parts shall be placed on service report, but shall
be billed as a no charge item on Contractor's invoice. All parts replaced, shall be
new/exchange replacement parts, as suggested by the original manufacturer.
If performance does not conform to the contract specification, the Government
will require the Contractor to perform the service again to conform to
contract specifications, at no increase in the contract cost. When the defects in
service cannot be corrected by re-performance, the Government may:
Require the Contractor to call in a certified technical expert on the listed instrument.
Specifically require the Contractor to call in a senior technical engineer from the manufacturer.
Repeated malfunctions of any systems component or subsection for more than five
times within one month shall not be tolerated. The Contractor is responsible for
repairing all faulty equipment and identifying the nature and cause if repeated
failures. The Contractor in this situation shall either:
Make repairs such that the identified problems are corrected to the extent that repeated failures are eliminated.
Prove that the equipment is not faulty and that an external cause for the problem exists with recommendations for resolution.
6. EQUIPMENT DESCRIPTION:
Equipment Identifiers
Equipment
EE 75977 SN: 2009
SYMBIA T16
If this is a service your business can provide, respond to this Sources Sought by Friday, November 29, 2024; 3:00 P.M. Eastern time. Responses should include: (1) Business Name and Address (2) Point of Contact Name, Phone Number, and E-mail Address (3) SAM UEI and NAICS code (4) Business Size SMALL or LARGE (5) Type of Business SERVICE DISABLED VETERAN OWNED, VETERAN OWNED, 8A, HUBZone, WOMEN-OWNED etc. (6) Applicable GSA/FSS Contract Number (if applies). All businesses eligible to provide this service are encouraged to reply as this information may be used to determine potential set-asides for the above noted service. E-mail responses to: Alexis Alexander, at Alexis.Alexander1@va.gov. (Note: This is NOT A REQUEST FOR QUOTE or an announcement of a solicitation.)
- VISN 10 CONSOLIDATED CONTRACTING 4100 WEST 3RD STREET
- DAYTON , OH 45428
- USA
- Alexis S. Alexander
- Alexis.Alexander1@va.gov
- Phone Number 734-222-7169
- Nov 22, 2024 12:38 am ESTSources Sought (Original)
Related Document
| Dec 3, 2024 | [Presolicitation (Original)] J065--Symbia T16 PM Service Agreement John D. Dingell VA Medical Center |
| Dec 10, 2024 | [Solicitation (Original)] J065--Symbia T16 PM Service Agreement John D. Dingell VA Medical Center |
| Dec 18, 2024 | [Solicitation (Updated)] J065--Symbia T16 PM Service Agreement John D. Dingell VA Medical Center |
See Also
*NOTICE ONLY - 2026-2027 M7B Agreement Bedford Res... Status: Open RFI 125532 Virginia
State Government of Virginia
Due by 2/12/2027
Virginia Lottery iLottery System, Services, and Su... Status: Open RFP 123127-1 Virginia State
State Government of Virginia
Due by 9/23/2026
Plumbing and Heating Maintenance and Repair Servic... Status: Open IFB 126991 Alexandria City
State Government of Virginia
Due by 10/07/2026
Follow Cleaning of Sanitation Tanks Active Contract Opportunity Notice ID N4215826Q0035 Related Notice
DEPT OF DEFENSE
Due by 9/28/2026