6530--EQUIPMENT - GULDMANN PATIENT CEILING LIFT SYSTEMS BRAND NAME ONLY
| Agency: | VETERANS AFFAIRS, DEPARTMENT OF |
|---|---|
| State: | New York |
| Type of Government: | Federal |
| FSC Category: |
|
| NAICS Category: |
|
| Posted Date: | May 6, 2026 |
| Due Date: | May 15, 2026 |
| Solicitation No: | 36C24226Q0596 |
| Original Source: | Please Login to View Page |
| Contact information: | Please Login to View Page |
| Bid Documents: | Please Login to View Page |
Description
APEX Accelerators are an official government contracting resource for small businesses. Find your local APEX Accelerator (opens in new window) for free government expertise related to contract opportunities.
APEX Accelerators are funded in part through a cooperative agreement with the Department of Defense.
The APEX Accelerators program was formerly known as the Procurement Technical Assistance Program (opens in new window) (PTAP).
- Contract Opportunity Type: Sources Sought (Original)
- Original Published Date: May 06, 2026 10:06 am EDT
- Original Response Date: May 15, 2026 03:00 pm EDT
- Inactive Policy: Manual
- Original Inactive Date: Aug 13, 2026
-
Initiative:
- None
- Original Set Aside:
- Product Service Code: 6530 - HOSPITAL FURNITURE, EQUIPMENT, UTENSILS, AND SUPPLIES
-
NAICS Code:
- 339113 - Surgical Appliance and Supplies Manufacturing
-
Place of Performance:
James J Peters VA Medical Center Bronx , NY 10468USA
The NAICS Code is 339113; size standard is 800 employees; PSC is 6530.
Unrestricted no set-aside
Responses to this notice shall include the following: (a) company name  b) address (c) socio-economic status, (d) point of contact (e) phone, fax, and email (f) SAM UEI number (g) Cage Code (h) Tax ID Number (i) and must  provide a capability statement that addresses the organizations qualifications and ability to perform as a contractor and provide the commodities described below. Â
BRAND NAME ONLY GULDMANN, INC.
OBJECTIVE:
James J. Peters VA Medical Center is replacing their outdated patient lifts that have exceeded their end-of-life. This project involves the deinstallation of the old units, procurement, installation, inspection, and staff training on the Guldmann GH3+ lifts, in coordination with Safe Patient Handling (SPH), Facilities Management Service (FMS), and Healthcare Technology Management (HTM).
STANDARDS AND SPECIFICATIONS:
Ceiling Lifts shall meet the latest versions of NFPA 13, NFPA 101, NFPA 70, NFPA 99, and all other applicable federal, state and local codes. Appropriate documentation should be provided.
Ceiling Lifts shall comply with ISO 10535: Hoists for the transfer of disabled persons Requirements and test methods .
Ceiling lifts must include a 2-point hangar bar, and manufacturers must allow and warrant the use of their lifts with the full range of manufacturer loop slings.
The lifts shall be mounted (using ceiling support as determined by building structure and room covering layout needs) to provide lifting capability for the maximum area possible within each room which means wall to wall.
The contractor shall offer both traverse and straight track options for the ceiling mounted lifts.
Lifts must be capable of picking patients up off the floor.
Scales must be internal; with patient weight displayed on LCD screen located on the hand control and must not decrease the lifting height of the system.
Hoist must not have an external scale attachment.
The lift hand control must be attached to the lift in a way that allows it not to be lost.
Ceiling lift motors shall be capable of lifting at least 600 pounds for standard lifts and 880 pounds for bariatric lifts, as specified.
Ceiling lifts vertical motion must be motorized.
Motors shall be GFCI protected.
Lift must charge automatically anywhere on the track via continuous AC charging.
Ceiling lifts must include safeguards to prevent patients from being trapped or hurt, such as emergency stop buttons or emergency lowering devices in case of power failure. Use of emergency lowering devices must not require tools or ladders and easily accessible to patient care personnel (i.e. Nursing).
Contractor shall not interfere with operations of existing sprinkler heads, lights, HVAC grills, curtain tracks, IV tracks, televisions, permanent furniture, or other permanent structures.
Contractor shall not breach any firewalls during the installation of these lifts. Any fire barrier that must be penetrated must be discussed with VA safety personnel and must immediately have fire barrier repaired. An Above Ceiling Permit must be filled out before ceiling work, and the COR will check the ceiling after ceiling work, to make sure any fire barrier penetration was repaired.
Each room where lifts are installed shall be returned to its original condition, excluding the newly installed ceiling lift. Holes in the wall or ceiling tile must be patched with appropriate materials.
Ceilings must not be slotted exposing space above ceiling to the room below.
Rail system and Patient Lift Motor must be provided by the same manufacturer.
Place of Delivery:
James J Peters VA Medical Center
130 W Kingsbridge Rd
Bronx, NY 10468
The Government is not obligated to, nor will it pay for or reimburse any costs associated with responding to this Sources Sought request. This notice shall not be construed as a commitment by the Government to issue a solicitation or ultimately award a contract, nor does it restrict the Government to a particular acquisition approach. The Government will in no way be bound to this information if any solicitation is issued. Notice to Potential Offerors: All Offerors who provide goods or services to the United States Federal Government must be registered in the System for Award Management (SAM located on the web at www.sam.gov). It is desirable that any Offeror to have completed their business Online Representations and Certifications Application (ORCA). Additionally, all Service-Disabled Veteran Owned Small Businesses or Veteran Owned Small Businesses who respond to a solicitation on this project must be registered and certified with the Small Business Administration (SBA). Requests for information and resultant responses shall be submitted directly to the contracting officer,
Response due no later than 15:00 Eastern Standard Time on May 15, 2026 to Paul Speedling, Contracting Officer at Paul.Speedling@va.gov.
- 113 HOLLAND AVE
- ALBANY , NY 12208
- USA
- Paul Speedling
- Paul.Speedling@va.gov
- Phone Number 914-737-4400 x202107
- May 06, 2026 10:06 am EDTSources Sought (Original)
Related Document
| Aug 12, 2026 | [Presolicitation (Original)] 6530--EQUIPMENT - GULDMANN PATIENT CEILING LIFT SYSTEMS |
| Aug 13, 2026 | [Combined Synopsis/Solicitation (Updated)] 6530--EQUIPMENT - PATIENT CEILING LIFT SYSTEMS |
| Aug 14, 2026 | [Combined Synopsis/Solicitation (Updated)] 6530--EQUIPMENT - PATIENT CEILING LIFT SYSTEMS |
| Sep 11, 2026 | [Award Notice (Original)] 6530--EQUIPMENT - PATIENT CEILING LIFT SYSTEMS |
See Also
Title: PV279ACON New York City Center Chiller Replacement Procurement Section Design and Construction
The City of New York
Due by 9/17/2026
32 Title: Spine System and Related Items CR#: 2138145 Agency: State University of
The New York State Contract Reporter
15 Title: Case Medical Rigid Sterilization Containers, Lids, Baskets. CR#: 2138673 Agency: State
The New York State Contract Reporter
Due by 9/24/2026
15 Title: SQUID Liquid Embolic Agent and related items manufactured by Balt USA
The New York State Contract Reporter
Due by 9/25/2026