Dynamic Mechanical Analysis Fatigue Testing System

Agency: Southwest Missouri State University
State: Missouri
Type of Government: State & Local
NAICS Category:
  • 541330 - Engineering Services
  • 541380 - Testing Laboratories
Posted Date: Mar 11, 2026
Due Date: Mar 26, 2026
Solicitation No: 8458-26
Original Source: Please Login to View Page
Contact information: Please Login to View Page
Bid Documents: Please Login to View Page
Solicitation Number Posted Date Return Date Description Buyer Name
8458-26 03-11-2026 03-26-2026 Dynamic Mechanical Analysis Fatigue Testing System

Attachment Preview

REQUEST FOR QUOTATION NUMBER 8458-26
FROM: The Board of Governors of Missouri State University
Procurement Services
901 South National Avenue
Springfield, Missouri 65897
Telephone: (417) 836-4414
Fax: (417) 836-6583
TO: (Provide company name and address below.)
Date:
March 11, 2026 Page 1 of 7
Bid Close Date and Time:
3:00 p.m. (CST) March 26, 2026
ALL BIDS MUST BE PREPARED
TO INCLUDE FREIGHT COSTS
TO SPRINGFIELD, MISSOURI
Buyer: Tanya Smith for Missouri State University
_____________________________________________________________________________________________________________________________________________
Missouri State University, as a state supported educational institution, is exempt from the payment of sales/use taxes pursuant to RSMo 144.040 and 144.615
DESCRIPTION/SPECIFICATIONS
RFQ 8458-26 Dynamic Mechanical Analysis/Fatigue Testing
System
Sealed bids are hereby requested for the following item(s).
General specifications and requirements are provided as part of this
RFQ. When submitting sealed bid, mark in lower left corner of
envelope: Sealed RFQ 8458-26.
All quoted prices must be firm and fixed and must include all
required components and shipping/delivery costs FOB Destination
to Missouri State University in Springfield, Missouri. Where
applicable, all quoted prices must be presented as Delivery Duty
Paid (DDP) to include Customs Fees, Freight/Shipping,
Import/Export Duties, and Taxes.
Faxed or e-mail bids will NOT be accepted.
THE BIDDER MUST SIGN AND RETURN THIS
DOCUMENT AS PART OF HIS/HER BID.
Unless noted by the bidder, there shall be no restrictions on the
number or quantity of an item that may be ordered.
If the requested quantity is not a standard package quantity, the
bidder should quote on both the requested quantity and the
standard package quantity, clearly identifying each.
If alternate items are offered, the bidder must provide descriptive
literature and/or information for each alternate offered.
Additional costs for packing, handling, etc., not shown on this
quotation will not be paid by the University.
See the “TERMS AND CONDITIONS OF REQUEST FOR
PROPOSAL/QUOTATION (RFP/RFQ)” attached hereto for
additional requirements.
DELIVERY_______DAYS ARO OR STATE EARLIEST____________________
PRICES EFFECTIVE FOR _______ DAYS OR UNTIL______________________
THE BIDDER HEREBY AGREES TO FURNISH THE ITEMS AND/OR
SERVICES SPECIFIED HEREIN, AT THE PRICES QUOTED, PURSUANT
TO ALL REQUIREMENTS AND SPECIFICATIONS OF THIS REQUEST
FOR QUOTATION.
AUTHORIZED SIGNATURE: ______________________________________
Please type or print name: _____________________________________________
TITLE____________________DATE_______________ PHONE NUMBER______________________
EMAIL ADDRESS_____________________________________________________________________
COMPANY NAME: __________________________________________________
RFQ 8458-26
MISSOURI STATE UNIVERSITY
Page 2 of 7
Specifications for DMA System:
General
1 The Jordan Valley Innovation Center seeks proposals for a servo-hydraulic fatigue testing/dynamic mechanical
analysis system capable of high-temperature cyclic and static mechanical testing of composite and ceramic materials
in accordance with:
1.1 ASTM C1341 (Flexural Properties of Continuous Fiber-Reinforced Advanced Ceramics at Elevated Temperature)
1.2 ASTM C1499 (Monotonic Equibiaxial Flexural Strength of Advanced Ceramics at Ambient Temperature)
2 System shall be configurable to support a wide range of additional ASTM standards including tension, compression,
flexural fatigue, etc. provided later acquisition of applicable fixturing and environmental solutions.
3 The system shall include all components necessary for complete system operation including load frame, control
electronics, pump, cables, plumbing and all necessary fittings and connectors. The system shall also include test
accessories with the features and capabilities described herein.
4 The components of the test system are to be new. No demonstration or used equipment will be acceptable.
Vendor Response
1 Vendor shall be willing to install the system in two or more phases if necessary to expedite delivery of the system. The
load frame and all components and accessories necessary to operate it shall be delivered as soon as manufacturing and
scheduling allow. The load unit shall include a high-temp furnace which the vendor shall agree to install during a later
appointment should the manufacturing time of the furnace significantly exceed that of the load frame.
System Load Unit
Load Frame
1 The system load frame shall be designed for static and dynamic testing and meet the following specifications:
1.1 Maximum dynamic load capacity: 25-30 kN
1.2 Dynamic load cell capacity: 2 kN
1.3 Stroke length: ±50 mm
1.4 Force accuracy: ±1.0% or better at 1/500 full scale
1.5 Hydraulic power unit suitable for continuous fatigue testing
1.6 Hydraulic line sets included
Load Cells
2 Dynamic load cell: 2kN
2.1 Include load cell adapter plate
2.2 Include calibration accessories if applicable
Control and Data Acquisition
3 The system shall include a digital controller and software capable of:
3.1 Load control
3.2 Displacement control
3.3 Static and dynamic test modes
3.4 User-defined waveform generation (sine, triangle, ramp, etc.)
3.5 Resonance tracking capability
3.6 Real-time data acquisition
RFQ 8458-26
MISSOURI STATE UNIVERSITY
Page 3 of 7
Hydraulic Performance Package
Hydraulic Power Unit
1 The hydraulic power unit (HPU) or pump shall supply 40 L/m. In addition, the HPU shall include pressure-control for
operating the HPU lower pressures.
2 Hydraulic power unit shall be designed to integrate with a building-wide chilled water/glycol loop for cooling.
High Temperature Capabilities
Furnace
1 Furnace shall have an operating temperature range of 1000°-1200°C with preference given to a system able to achieve
1200°C. Ability to maintain stated temperature for up to thirty days.
2 Furnace shall have a hot zone capacious enough to accommodate a 4-point-1/4 bend sample with dimensions 3mm
thick by 9mm wide by 60mm long and appropriate fixturing as described in ASTM C1341-13
3 Shall have ports to accommodate an extensometer as applicable
4 Shall have inlet/outlet ports for purge gases.
Fixturing
1 Quoted system shall include fixtures for both standards indicated by this RFQ
2 Push/pull rods and fixturing solution for ASTM C1341-13 shall be rated for the maximum operating temperature
capability of the system.
2.1 Preferred that room-temperature fixturing is compatible with a Shimadzu AGX-V2
3 Fixturing for higher force capacities may be included as optional accessories.
Extensometer
1 Shall include a contact or optical extensometer:
1.1 Shall be configurable to meet the deflection measurement required by ASTM C1341-13
1.2 Shall have measurement capability that meets the stroke range of the load frame
1.3 Shall be rated for the maximum operating temperature capability of the system (i.e. if you offer a system capable of
1000°C the extensometer must also be capable of operating at 1000°C).
1.4 Shall include calibration/verification post install
RFQ 8458-26
MISSOURI STATE UNIVERSITY
Page 4 of 7
Pricing
State a firm, fixed total price to provide a Dynamic Mechanical Analysis/Fatigue Testing System:
$____________ firm, fixed total price
Specify brand and model number of item bid: _______________________________________________
Bid Submission Information and Requirements
1. The bidder must provide a detailed, itemized description of the proposed equipment to include all related
components and accessories. Such description should also include a written narrative which identifies the
relationship of each component or accessory to the specifications contained herein.
2. The bidder should provide any other relevant information that should be considered by Missouri State University
in evaluating the proposed equipment. The bidder is advised that the University is under no obligation to seek
out information regarding the proposed equipment if such information is not included as part of the bidder’s
response.
3. All quoted prices must be firm and fixed and must include all required components and shipping/delivery costs
FOB Destination to Missouri State University in Springfield, Missouri.
4. Bid responses must be received in the Procurement Services Office of Missouri State University on or before 3:00
p.m. (CST) on March 26, 2026 in order to be considered for award. Responses received after the exact time on
the date specified will not be opened and will not be evaluated regardless of the reason(s) or mitigating
circumstances related to its lateness or degree of lateness.
Bidders are reminded that SEALED BIDS are required. Bid responses may NOT be faxed or e-mailed.
Please attach your formal quote to your sealed RFQ response.
Mail bid responses to:
The Board of Governors of Missouri State University
Procurement Services
Attention: Tanya Smith
901 South National Avenue
Springfield, Missouri 65897
Mark in lower left corner of envelope: Sealed RFQ 8458-26
RFQ 8458-26
MISSOURI STATE UNIVERSITY
Page 5 of 7
TERMS AND CONDITIONS, REQUEST FOR PROPOSAL/QUOTATION (RFP/RFQ)
I. TERMINOLOGY/DEFINITIONS
1. Bids and Bidders means the same as proposals and offerors.
2. Closing Date and Time or Date and Time Returnable means the deadline required for the physical receipt of bids by the Procurement Office.
3. Must and Shall mean that a certain feature, component, or action is mandatory. Failure to comply will result in rejection of the proposal.
4. Request for Proposal (RFP) or Request for Quotation (RFQ) means the procurement document issued by the Procurement Office to potential bidders for the
purchase of equipment, supplies, and/or services as described in the RFP/RFQ document. The definition includes these Terms and Conditions as well as any
other Attachments and Amendments to the RFP/RFQ.
II. OPEN COMPETITION/RFP/RFQ DOCUMENT
1. It is the offeror’s responsibility to ask questions, request changes or clarification, or otherwise advise the Procurement Office if any language, specifications,
or requirements of the RFP/RFQ appear to inadvertently restrict or limit the requirements of the RFP/RFQ to a single source. Any and all communication
from bidders regarding the RFP/RFQ and the procurement process must be directed to the Buyer identified on the first page of the RFP/RFQ. Such
communication should be received at least five calendar days prior to the official Proposal Close Date.
2. The Procurement Office reserves the right to officially modify or cancel the RFP/RFQ after issuance. Bidders will be notified of any such modification or
cancellation by issuance of an addendum from the Procurement Office.
III. CONFLICT OF INTEREST
The contractor hereby covenants that at the time of the submission of the proposal the contractor has no other contractual relationships which would create any
actual or perceived conflict of interest. The contractor further agrees that during the term of the contract neither the contractor nor any of its employees shall
acquire any other contractual relationships which could create such a conflict.
IV. PREPARATION OF BIDS/PROPOSALS
1. Failure to carefully examine the RFP/RFQ and all terms and conditions related thereto will be at the Offeror’s risk.
2. Unless otherwise specifically stated in the RFP/RFQ, all specifications and requirements constitute minimum requirements. All bids must meet or exceed the
stated specifications and requirements.
3. Bids lacking any written indication of intent to offer an alternate brand or to take an exception shall be considered to be in full compliance with the specifications
and requirements of the RFP/RFQ.
4. All equipment and supplies offered must be new and of current production and available for marketing by the manufacturer unless the RFP/RFQ clearly
specifies that used, reconditioned, or remanufactured equipment and supplies may be offered.
5. Prices quoted by the offeror must remain valid for a minimum of ninety (90) days from the date of proposal opening. If the proposal is accepted, prices shall
remain firm for the specified period.
V. SUBMISSION OF BIDS/PROPOSALS
1. Bids must be signed by a duly authorized representative of the offeror’s organization and must contain all necessary information in the manner required by the
RFP/RFQ.
2. A bid may only be modified or withdrawn (1) by signed, written notice received by the Procurement Office prior to the Proposal Close Date and Time or (2)
in person by the offeror or its authorized representative, provided proper identification is presented before the Bid Close Date and Time. Telephone requests
to withdraw or modify a proposal will not be honored.
3. Bidders must, as a minimum, sign and return the RFP/RFQ cover page or, if applicable, the cover page of the last amendment thereto in order to constitute
acceptance by the offeror of all RFP/RFQ terms and conditions. Failure to do so will result in rejection of the proposal unless the offeror’s full compliance
with those documents is indicated elsewhere within the offeror’s response.
VI. QUOTATION/PROPOSAL OPENING
1. All bids will be opened at 3:00 p.m. on the date indicated. Prices will not be made available at that time and will not be made public until the evaluation is
complete and an award is made.
2. It is the offeror’s sole responsibility to ensure that the proposal is delivered to the Procurement Office by the Proposal Close Date and Time. If the Procurement
Office is officially closed on the Proposal Close Date, bids will be accepted until 3:00 p.m. the next official work day and will be opened at that time.
3. Bids which are not received by the Procurement Office prior to the Proposal Close Date and Time will be considered late, regardless of the degree of lateness
or the reason related thereto. Late bids will not be opened.
This page summarizes the opportunity, including an overview and a preview of the attached documents.
* Disclaimer: This website provides information about bids, requests for proposals (RFPs), or requests for qualifications (RFQs) for convenience only and does not serve as an official public notice. Individuals who wish to respond to or inquire about bids, RFPs, or RFQs should contact the relevant government department directly.

Sign-up for a Free Trial, Government Bid Alerts

With Free Trial, you can:

You will have a full access to bids, website, and receive daily bid report via email and web.

Try One Week FREE Now

See Also

Follow Replacement Radiosondes for the National Logistics Support Center in Grandview, MO. Active

COMMERCE, DEPARTMENT OF

Bid Due: 6/15/2026

Bid Information Type Quote (informal) Status Issued Number 26-103 (Medical Emergency Kit Supplies)

Springfield Public Schools

Bid Due: 6/16/2026

Number: 677092 Title: Must be OEM items only and meet all Bi-State specifications.

St. Louis region

Bid Due: 6/08/2026

Number: 677081 Title: Must be OEM items only and meet all Bi-State specifications.A

St. Louis region

Bid Due: 6/08/2026