Automated Liquid Handling Platform
| Agency: | HEALTH AND HUMAN SERVICES, DEPARTMENT OF |
|---|---|
| State: | Maryland |
| Type of Government: | Federal |
| FSC Category: |
|
| NAICS Category: |
|
| Set Aside: | No Set aside used |
| Posted Date: | Sep 17, 2026 |
| Due Date: | Sep 22, 2026 |
| Solicitation No: | 75F40126Q00423 |
| Original Source: | Please Login to View Page |
| Contact information: | Please Login to View Page |
| Bid Documents: | Please Login to View Page |
Description
APEX Accelerators are an official government contracting resource for small businesses. Find your local APEX Accelerator (opens in new window) for free government expertise related to contract opportunities.
APEX Accelerators are funded in part through a cooperative agreement with the Department of Defense.
The APEX Accelerators program was formerly known as the Procurement Technical Assistance Program (opens in new window) (PTAP).
- Contract Opportunity Type: Presolicitation (Original)
- Original Published Date: Sep 17, 2026 04:30 pm EDT
- Original Response Date: Sep 22, 2026 11:00 am EDT
- Inactive Policy: Manual
- Original Inactive Date: Sep 23, 2026
-
Initiative:
- None
- Original Set Aside: No Set aside used
- Product Service Code: 6640 - LABORATORY EQUIPMENT AND SUPPLIES
-
NAICS Code:
- 334516 - Analytical Laboratory Instrument Manufacturing
-
Place of Performance:
Silver Spring , MD 20993USA
All equipment shall be newly manufactured. It shall not be used, refurbished, reconditioned, remanufactured, or previously used for demonstration.
Salient characteristics and any single-source need
The equipment shall meet the following salient characteristics: a benchtop automated liquid handling platform, user-friendly and easy to operate and clean, of space-saving and energy-saving design, capable of fast automatic pipetting of multiple solutions and of serial dilutions across or down the plate; able to accommodate variable liquid volumes; able to pipette into single rows, single columns, or individual wells as selected by the user in a 96-well or 384-well format, with the well format interchangeable based on user requirements; customizable to the laboratory's assay needs; equipped with programmable operating temperatures permitting use across a range of temperatures; and usable within the containment enclosure identified in the minimum technical requirements below to permit handling of sterile or hazardous materials.
Based on our research, only a specific make and model meets our needs, the Agilent Technologies Bravo Automated Liquid Handling Platform. It is the only equipment in its class that can meet our minimum needs as listed in the specifications. Other machines do not dispense low volumes, are much larger than this machine and do not fit within a biosafety cabinet.
The equipment shall meet the following minimum requirements:
-
The platform shall be benchtop in format and shall fit and operate within a biosafety cabinet of dimensions of 46in W X 24in D X 30in H that will support 120 lbs without disrupting the enclosure airflow.
-
The platform shall perform serial dilutions either across the plate or down the plate using standard disposable tip heads.
-
The platform shall use 96-channel or 384-channel pipetting heads that are interchangeable, with independent tip pickup capability.
-
The platform shall provide a minimum 15-position deck capacity.
-
The platform shall dispense from 0.1 to 200 microliters. Dispensing accuracy and precision shall be no worse than 5% deviation.
-
The platform shall be configurable to support shaking, heating, cooling, and filtration. These configurations are not required to be delivered with the platform; however, the platform shall be capable of accepting these configurations at a later date, if desired.
-
The platform shall include a controller and software to generate custom protocols.
-
The platform shall include a dedicated computer to control the machine.
-
The platform shall be compatible with standard format microplates (96-well and 384-well).
-
The platform shall operate under ambient conditions of 5 to 40 degrees Celsius and 10 to 90 percent relative humidity, non-condensing. No deck or plate temperature-control range is required.
-
Compliance with all applicable Federal, state, and local safety standards.
- 11601 Landsdown St Floor 13
- Rockville , MD 20852
- USA
- Reagan Davis
- reagan.davis@fda.hhs.gov
- Phone Number 3017960910
- Sep 17, 2026 04:30 pm EDTPresolicitation (Original)
Related Document
| Sep 22, 2026 | [Presolicitation (Updated)] Automated Liquid Handling Platform |
See Also
Follow Instruments Upgrade - Nitrogen & Hydrogen Generators Active Contract Opportunity Notice ID
DEPT OF DEFENSE
Due by 9/29/2027
Follow USCG Yard: Sensors and Meters Active Contract Opportunity Notice ID 70Z04027Q60002Y00 Related
HOMELAND SECURITY, DEPARTMENT OF
Due by 11/09/2026
Follow Sources Sought for (ICP-OES) Spectrometer System with an Autosampler Active Contract Opportunity
DEPT OF DEFENSE
Due by 12/22/2026
Follow Business Contact and Firmographics Data Active Contract Opportunity Notice ID 041126-26-2218B Related
COMMERCE, DEPARTMENT OF
Due by 11/19/2026